Instructions for use Primus Infinity Empowered SW 4.5n 139
Alarms
To restore individual alarm limits for MV and PAW:
z see "Setting alarm limits" on page 171.
To restore all default alarm limits:
z see "Operation" on page 92.
Setting the alarm tone
The volume of the alarm tone can be set in the con-
figuration menu, see page 167.
Alarms in the Standby mode
All technical alarms, e.g., failure of device compo-
nents and a number of special operating states, are
also indicated by Primus Infinity Empowered when
in Standby.
A corresponding message appears in the alarms
field, but without any acoustic warning.
Volume Mode,
Volume AF,
Pressure Mode,
Pressure Support
P
AW
upper alarm limit
[hPa (cmH
2O)]
PEAK +5
or
PLAT +10,
the greater value applies
P
AW
lower alarm limit
[hPa (cmH
2O)]
0.6 x (PLAT – PEEP)
+ PEEP – 1,
but at least 3