Maintenance
242
Instructions for use Primus Infinity Empowered SW 4.5n
Technical customers documentation according to
IEC/EN 60601 is available upon request.
Repair
Dräger recommends that all repairs are carried out
by DrägerService and that only authentic Dräger
repair parts are used.
Man.Spont.-Auto-
matic reversing dia-
phragm
Every 2 years Replace Service personnel
Bacterial/viral filter,
patient gas module
Every 2 years Replace Service personnel
Bacterial/viral filter,
ventilation and gas
controller
Every 2 years Replace Service personnel
O-rings between
valve plate and dia-
phragm cover of
breathing system
Every 2 years Replace Service personnel
Lower diaphragm of
ventilator unit
+ O-ring
Every 3 years Replace Service personnel
Lead gel battery in
UPS
Every 3 years Replace Service personnel
Or when message BAT-
TERY LOW is dis-
played.
Replace Experts
Component Interval Measure Personnel responsible