Instructions for use Primus Infinity Empowered SW 4.5n 219
Cleaning, disinfection and sterilization
Reusable absorber
1 Turn the absorber counterclockwise and pull it
down (C).
2 Empty out the soda lime in accordance with the
instructions for use of the absorber, see
page 123.
3 Remove the insert from the absorber (D). The
inner and outer sealing rings remain on the ab-
sorber insert.
4 Prepare the absorber for conditioning in a
cleaning and disinfecting machine.
Removing the breathing system
1 Loosen the three sealing screws (A) on the ven-
tilator a quarter turn counterclockwise using the
key supplied.
049050
C
D
NOTE
Before removing the breathing system, allow it to
cool 5 minutes if the anesthesia machine has just
been used. The surface may otherwise be hot to
the touch.
051
A
A
A