Operation
92
Instructions for use Primus Infinity Empowered SW 4.5n
Operation
Safety Information
Loading therapy-related data from
Infinity ID breathing circuits
Primus Infinity Empowered supports automatic
configuration of the workplace by transfer of ther-
apy-related data via Infinity ID breathing circuits
from and to other devices supporting this function-
ality. Transfer of the following data is supported:
Ventilation settings and ventilation therapy data
Fresh-gas settings
Alarm limits
Patient data
Ventilation Settings
Ventilation settings are automatically transferred.
Ventilation mode settings, which are different from
those offered in Primus Infinity Empowered, are au-
tomatically transferred into that mode offering the
most similar respiratory support.
WARNING
Risk of electric shock
Touching the patient and electrical device
contacts could result in an electric shock.
Do not touch the patient and the electrical de-
vice contacts at the same time.
NOTE
When connecting an Infinity ID breathing circuit to
the breathing system, no inspiratory or expiratory
microbial filters should be used.