For your safety and that of your patients
14
Instructions for use Primus Infinity Empowered SW 4.5n
Functional safety
The essential performance consists in:
Supplying the anesthesia workstation with O
2
If the O2 supply (central supply or gas cylinder)
fails, an alarm is issued.
Supply of the patient with adequately oxygen-
ated breathing gas
If the breathing gas contains insufficient levels
of O2, an alarm is issued.
Patients are not supplied with excessively high
anesthetic gas concentrations
If excessively high anesthetic gas concentra-
tions are delivered, an alarm is issued.
Monitoring the airway pressure and the expira-
tory minute volume
Alarms are issued depending on the set alarm
limits.