Instructions for use Primus Infinity Empowered SW 4.5n 13
For your safety and that of your patients
CAUTION
Risk of patient injury
An incorrect diagnosis or misinterpretation of
measured values, or other parameters, may en-
danger the patient.
Do not base therapy decisions on individual
measured values or monitoring parameters only.
WARNING
Risk of patient injury
If ventilation of the patient is no longer en-
sured due to an obvious fault in the equip-
ment, the patient must immediately be venti-
lated with a manual resuscitator.
Always keep an emergency manual breathing
bag at hand.
WARNING
Risk of burns
Conductive breathing hoses or face masks
may cause burns during HF surgery.
Do not use these types of hoses and masks in
combination with HF surgery.
CAUTION
Risk of mechanical failure
The shock and vibrations caused by transporta-
tion may lead to a mechanical failure. The appli-
cation of a wall or ceiling mount is designated for
buildings.
Do not use the anesthesia machine for mobile fa-
cilities such as ambulances, helicopters or ships.
CAUTION
Risk of crushing
Movable parts and attached parts can lead to
crushing injuries. Pay special attention to edges,
movable parts, and corners when working with the
following parts:
– Drawers
– Ventilator module
–Doors
– Writing table
– Swivel arms for mounted devices
– Accessories such as gas cylinders, vaporiz-
ers, CLIC absorbers, and CLIC adapters
CAUTION
Risk of device failure
Compressed gas supply (central supply or cylin-
der): To avoid damaging the device(s) attached to
a gas supply, use only medical gases. Pay partic-
ular attention to national and international stand-
ards regulating the use of medical gases.