When finishing oxygen therapy, make sure the flow
meter is completely closed:
z Turn the flow knob (E) clockwise until it can no
longer be turned.
Only then the oxygen flow is completely off.
Emergency resuscitator
Example: Dräger Resutator 2000
z The emergency resuscitator is present on the
device and its functionality has been checked.
Note the instructions for use.
Vaporizers
The vaporizers used must conform to the
ISO 8835-4 or ISO 80601-2-13 standard. If the in-
ternal patient-gas measurement module fails, an in-
dependent gas measurement system conforming
to ISO 21647 or ISO 80601-2-55 must be used.
The Vapor 2000 is shown and described here.
For the Dräger Interlock 2 system:
z Vaporizers are mounted in a level position and
seated securely on the mounts.
1 Locking levers (A) point to the left = locked po-
sition.
2 Check the sight glasses (B) and ensure an ad-
equate filling level.
3 Ensure that the control dial is set to 0 and button
(C) is engaged.
WARNING
Risk of fire
Cauterizing close to a source of oxygen can
lead to fire. Make sure that all connectors
(e.g., Y-piece, breathing hoses including the
breathing bag, breathing system, external
fresh-gas outlet, oxygen therapy, anesthetic
gas scavenging system) are tight so that oxy-
gen leaks cannot endanger the user or the pa-
tient.
CAUTION
Risk of inadequate pressure monitoring
The optional auxiliary outlets are not pressure
monitored.
Pressure monitoring must be ensured by the con-
nected device.
NOTE
Before operating the vaporizer, pay special atten-
tion to the instructions for use of the vaporizer be-
ing used. Note especially the vaporizer flow limits.