Cleaning, disinfection and sterilization
232
Instructions for use Primus Infinity Empowered SW 4.5n
9 Insert the absorber in the breathing system from
below (D) and turn clockwise as far as possible.
The disposable Drägersorb CLIC absorber can
also be used instead of the reusable absorber
described here. (See the Drägersorb CLIC in-
structions for use.)
If the breathing system is not to be used within the
next 24 hours:
z Only fill with the soda lime immediately before
use!
z If no Infinity ID CLIC absorber is used, reset the
soda lime change log to current date by press-
ing the soft key soda lime changed,
seepage88.
Note the instructions for use of the
Drägersorb 800 Plus or Drägersorb FREE soda
lime.
Installing the flexible arm (optional) and
breathing bag
1 Position the socket of the arm on the breathing
system and tighten it with the two knurled
screws (A).
2 Check that the arm is fixed securely!
078
D
WARNING
Risk of patient injury
The soda lime loses moisture. Generally, if the
moisture level falls below a minimum set
point, undesirable reactions occur, independ-
ent of the type of lime and the volatile anes-
thetic agent being used:
–reduced CO
2 absorption,
increased heat build-up in the absorber
and thus, an increased breathing gas tem-
perature,
formation of CO,
absorption and/or decomposition of the in-
halation anesthetic.
These reactions could pose a danger to the
patient.
If using dry gases, only briefly flush the an-
esthesia system and only if necessary.
076