Instructions for use Primus Infinity Empowered SW 4.5n 235
Cleaning, disinfection and sterilization
5 Connect the gray transfer hose to the scaveng-
ing nozzles (D) on the Primus Infinity
Empowered and on the AGS (E).
6 Connect the scavenging hose to the scaveng-
ing nozzle of the AGS (F).
7 Connect the scavenging hose to the scaveng-
ing connector (G).
8 Make sure the second port of the scavenging
system (H) is sealed with the screw plug.
9 Connect the scavenging connector (G) to the
terminal unit of the disposal system. The opera-
tion indicator of the terminal unit is green.
10 The AGS is functioning when the float in the
flow tube is between the two marks.
Note the instructions for use of the anesthetic gas
scavenging system AGS.
Connecting the secretion aspirator
1 Press the overflow protection float (A) into the
rising pipe until it clicks into place. The float
must move freely without falling out.
2 Insert the cover (B) of the secretion collecting
bottle into the silicon sleeve (C).
3 Grip the silicone sleeve of the bottle cover and
fit it on the bottle (D).