Cleaning, disinfection and sterilization
236
Instructions for use Primus Infinity Empowered SW 4.5n
4 Place the secretion collecting bottle in the inner
sleeve (E) and the rinsing bottle in the outer
sleeve (F).
When using the disposable VacuSmart con-
tainer:
z Place the VacuSmart container in the secre-
tion collecting bottle and press the sleeve
tightly into place.
5 Install the filter in the bottom of the endotracheal
aspiration system (G).
6 Connect the vacuum hose to the filter outlet of
the endotracheal aspiration system (H) and to
the thin port on the bottle cap.
7 Connect the suction hose to the thick port on
the bottle cap (I).
Observe the instructions for use of the endotra-
cheal aspiration system.
Before using on patients on again
1 Re-assemble all equipment, see "Assembly
and preparation" on page 51.
2 Check readiness for operation, see "Getting
started" on page 73.