Instructions for use Primus Infinity Empowered SW 4.5n 95
Operation
Setting ranges and factory settings
Setting fresh-gas concentrations
1 O2 concentration O2 % (A)
2 Fresh-gas flow flow L/min (B).
The fresh-gas settings can be changed before se-
lecting a ventilation mode. Fresh gas does not flow
in the Standby mode (soft keys = light green).
A corresponding text No fresh-gas delivery! (C) is
displayed. The fresh-gas flow is not enabled until a
ventilation mode has been started (soft keys = dark
green).
Setting ranges and factory settings
Parameter Setting
ranges
Factory
setting
age <1 to 120 years 40
weight 1 to 120 kg
(1 to 240 lbs)
– –
305
Standby
Standby
AB
C
Fresh-gas
parameters
Setting
ranges
Factory
setting
Carrier gas Air or N
2OAIR
O
2
[%]
21 to 100 for
carrier gas Air
25 to 100 for
carrier gas
N
2O
100
Fresh-gas flow
[L/min]
0.2 to 18 2