Application
18
Instructions for use Primus Infinity Empowered SW 4.5n
Indications and contraindications
Indications
Primus Infinity Empowered is intended for inhala-
tional anesthesia and/or patient ventilation in ac-
cordance with its intended use during surgical or di-
agnostic procedures.
Contraindications
For patients with suspected malignant hyperther-
mia: Do not use any volatile anesthetic agents or
Primus Infinity Empowered with residual concentra-
tions of these gases above 5 ppm.
Do not perform long term low-flow anesthesia on
patients with ketoacidosis or patients under the in-
fluence of alcohol. Otherwise, there is a risk of ac-
etone accumulation in the patient.
Further information on application
Environment of use
Primus is designed for use in areas in which thera-
peutic or diagnostic procedures can be performed.
Do not use Primus in the following environments:
Outside buildings
On intensive care units
During patient transport
In vehicles, airplanes, or helicopters
WARNING
Risk of explosion and fire
The medical device is not approved for use in
areas where oxygen concentrations greater
than 25 Vol%, or combustible or explosive gas
mixtures are likely to occur.
WARNING
Risk of device failure and/ or danger to patient
Magnetic fields may negatively influence the
proper function of the medical device, thus
endangering the patient or user.
The medical device must not be used in the vi-
cinity of magnetic resonance imagers (MRI,
NMR, NMI).