Instructions for use Primus Infinity Empowered SW 4.5n 61
Assembly and preparation
Caution when handling O2 cylinders
The cylinder valves must be opened/closed slowly
by hand. Do not use tools.
If a cylinder valve is leaky or difficult to operate, it
must be repaired by trained personnel.
Connecting the anesthetic gas scavenging system AGS
According to the particular requirements for an-
esthesia workstations, the use of an anesthetic gas
scavenging system is required.
1 Connect the gray transfer hose (A) to the noz-
zles on the Primus Infinity Empowered and on
the AGS.
2 Connect the scavenging hose to the scaveng-
ing nozzle (B) of the AGS.
3 Connect the scavenging hose to the scaveng-
ing connector (C).
4 Make sure the second connection to the scav-
enging system is sealed by a screw plug (D).
5 Connect the scavenging connector (C) to the
terminal unit of the anesthetic gas scavenging
system. The operation indicator of the terminal
unit is green.
6 The AGS is functioning when the float in the
flow tube is between the two marks.
Observe the instructions for use included with the
anesthetic gas scavenging system AGS.
WARNING
Risk of explosion
If the O
2 cylinder valves or O2 pressure reduc-
ing adapters is handled with oily or greasy fin-
gers/hands, the risk of explosion is eminent.
Do not oil or grease the O2 cylinder valves or
O2 pressure reducing adapters, and do not
handle with oily or greasy fingers.
WARNING
Risk of supply failure
If pressure reducers not having the required
sensors and measurement features are used
instead of Dräger pressure reducers, reserve
cylinders and their fill levels will not be cov-
ered by alarm and monitoring functionalities
during the power-on self test and operation.
Without this monitoring, it cannot be guaran-
teed that, in the event of a loss of the central
supply, the backup functionality of the reserve
gas cylinders will be available. If monitoring
for the remaining capacity of the backup gas
cylinders is not available, the user must take
other equivalent measures.