Instructions for use Primus Infinity Empowered SW 4.5n 17
Application
The following parameters can be
displayed as mini trends*
– Minute volume CO2 MV*CO2
–O2 Uptake
– PEEP, patient compliance C
PAT
The following parameters are displayed
as curves
– Airway pressure PAW
– Inspiratory and expiratory flow
– Inspiratory and expiratory concentration of O
2,
CO2, and anesthetic gas
Optional:
– Plethysmogram
– PAW-V loops and V-Flow loops
The following are displayed as bar
graphs
– Inspiratory tidal volume, expiratory tidal volume,
and leakage tidal volume
– Volumeter
–Pressure
– Econometer for indicating fresh-gas utilization
(optional)
Trends showing the measured values over time
and a logbook are also available.
Monitoring
by means of adjustable alarm limits which can au-
tomatically be adapted to the momentary ventila-
tion situation.
With monitoring for
– Airway pressure P
AW
– Expiratory minute volume MV
– Apnea
– Inspiratory and expiratory anesthetic gas con-
centration
– Detection of anesthetic gas mixtures (simulta-
neous detection of up to two anesthetic agents)
– Inspiratory O
2 and N2O concentrations
– Inspiratory and expiratory CO
2 concentrations
– Special alarm behavior in HLM mode
– Automatic agent alarm activation for multiples
of MAC (xMAC)
Optional:
– Oxygen saturation
–Pulse rate
*optional