Alarms
134
Instructions for use Primus Infinity Empowered SW 4.5n
Enabling/disabling CO2 alarms
The alarm limits for inCO2, etCO2, and CO2 apnea
monitoring can be disabled via the soft key CO2
alrm ON ->off. This key is effective in the following
operating modes:
Man.Spont.
Monitoring
Ext. Outlet
Disabling CO2 alarms
z Press the soft key CO2 alrm ON ->off (A).
The symbol appears beside the measured val-
ues for the end-expiratory and inspiratory CO
2 con-
centration (B).
Enabling the CO2 alarms
z Press the soft key CO
2 alrm OFF ->on (A) again.
Disabled CO
2 alarms are enabled automatically
when changing to another ventilation mode.
The alarms for etCO
2 and inCO2 can be
activated or deactivated in Standby for switching to
Man.Spont..
When the alarm limits are enabled the value is
adopted from the automatic ventilation mode,
see page 171.
CO
2 alarms can also be enabled and disabled
globally for all ventilation modes:
1 Press the soft key config. on the standard or
data screen. The submenu volumes/ alarms is
opened (C).
2 Select and confirm the column alarms on/off
(D) via the rotary knob.
3 Select and confirm the line CO
2 (E) via the ro-
tary knob.
4 Select and confirm on or off via the rotary knob.
336
Man.Spont.
etCO2
inCO2
B
A
337
CAUTION
Risk of inadequate monitoring
National and international standards require a
minimum monitoring with some alarm functions.
These standards may not be met if the alarm func-
tion of the etCO2 monitoring parameter is disa-
bled.
Only disable this monitoring parameter after con-
sulting national and/or international standards.
D
E
Volume
C