Instructions for use Primus Infinity Empowered SW 4.5n 133
Alarms
All apnea, apnea pressure, apnea flow, and apnea
CO
2 alarms are active after at least 15/30 seconds
depending on the set frequency/Freq
MIN.
All apnea and limit-based O
2, CO2, N2O, and agent
alarms are only active if a breath has already been
detected (AutoWakeUp).
inDes
[Vol%]
ON
ON
ON
ON
ON
1)
ON
1)
12.0
– –
inSev
[Vol%]
ON
ON
ON
ON
ON
1)
ON
1)
4.2
– –
PAW
[hPa
(cmH
2O)]
ON
ON
ON
ON
ON
OFF
ON
ON
40
8
APNEA PRES-
SURE
ON OFF OFF OFF 8
APNEA FLOW ON ON OFF OFF - - -
APNEA CO2 ON ON ON
2)
ON
2)
- - -
1) In Standby, these alarms can be configured ON or OFF for switching to Man. Spont., Ext. Outlet, and Monitoring. When
the alarm limits are set to ON the value is adopted from the automatic ventilation mode. The default value for this config-
uration is OFF.
2) In Man.Spont., Monitoring, and Ext. Outlet, the alarm is triggered after 65 seconds.
– The factory setting is outside the monitored range; the corresponding alarm limit is disabled.
Mode
Alarm
Volume, Volume AF,
Pressure,
Press. Support
Press.
Support
CPAP
Ext. Outlet Monitoring,
Man.Spont.
Factory setting