Technical data
250
Instructions for use Primus Infinity Empowered SW 4.5n
O2-Flush >35 L per minute
Standardized to ATPD conditions.
O
2 safety flow 0 to 12 L per minute
Standardized to STPD conditions.
Accuracy ±40 % or ±2 L per minute, the greater value applies
Optional outlets For connecting O2 flow meters and endotracheal suction.
Pressure corresponds to inlet pressure.
The outlets are not pressure monitored; this monitoring must
be ensured by the connected device.
Air outlet Maximum permitted flow: 70 L per minute
Standardized to STPD conditions.
O
2 outlet Maximum permitted flow: 20 L per minute
Standardized to STPD conditions.
Ventilator
Electronically controlled, electrically driven piston ventilator, fresh gas decoupled
Ventilation modes Volume Mode, Pressure Mode, Volume AF Mode (op-
tional), CPAP (optional) and Pressure Support (optional)
Settings
Pressure limitation P
MAX
In Volume Mode and Volume AF
Mode
(PEEP +10) to 70 hPa [(PEEP +10) to 70 cmH2O]
Accuracy ±10 % of set value or at least ±3 hPa (±3 cmH
2O)
Inspiration pressure P
INSP
In Pressure Mode (PEEP +5) to 70 hPa [(PEEP +5) to 70 cmH2O]
Accuracy ±10 % of set value or at least ±3 hPa (±3 cmH
2O)