Instructions for use Primus Infinity Empowered SW 4.5n 265
Technical data
Operating data
Operating voltage 100 to 240 V~, 50/60 Hz, 12.8 A max.
Power consumption
at 230 V∼
Standby (without auxiliary outlets) 0.8 A
Typical (without charging the internal
battery, without auxiliary outlets)
0.9 A
Maximum (with auxiliary outlets) 12.8 A
at 110 V
∼
Standby (without auxiliary outlets) 1.6 A
Typical (without charging the internal
battery, without auxiliary outlets)
1.8 A
Maximum (with auxiliary outlets) 12.8 A
Power input
Standby 180 W
Typical 200 W
Maximum (with power consumption
on auxiliary outlets)
2.5 kW
Internal battery
Type Lead-gel battery
Sealed, maintenance-free
Backup time with new and fully
charged battery (auxiliary outlets not
supplied)
At least 30 minutes
Maximum 90 minutes (depending on ventilation parameters)
Charging time
(to reach full power)
At least 10 hours
Charging power Maximum 70 W