Alarms
132
Instructions for use Primus Infinity Empowered SW 4.5n
Limit-based alarms activated in respective ventilation modes
Some alarms can be enabled and disabled individ-
ually in different operating modes.
WARNING
Risk of patient injury
Because anesthesia machines within one care
area might have different default alarm limit
configurations, make sure the preset alarm
limits are appropriate for the new patient. Also
make sure the alarm system has not been ren-
dered useless by setting the alarm limits to ex-
treme values or by their being disabled.
See "Configuring the default settings"
on page 166.
Mode
Alarm
Volume, Volume AF,
Pressure,
Press. Support
Press.
Support
CPAP
Ext. Outlet Monitoring,
Man.Spont.
Factory setting
SpO
2
[%]
ON
ON
ON
ON
ON
ON
ON
ON
– –
92
Pulse
[1/min]
ON
ON
ON
ON
ON
ON
ON
ON
120
50
etCO
2
[mmHg]
ON
ON
ON
ON
1)
1)
1)
1)
50
– –
inCO2
[mmHg]
ON ON
1) 1)
5
MV
[L/min]
ON
ON
ON
ON
OFF
OFF
1)
1)
12
3.0
inO2
[Vol%]
ON
ON
ON
ON
1)
ON
1)
ON
– –
20
inHal
[Vol%]
ON
ON
ON
ON
ON
1)
ON
1)
1.5
– –
inIso
[Vol%]
ON
ON
ON
ON
ON
1)
ON
1)
2.3
– –
inEnf
[Vol%]
ON
ON
ON
ON
ON
1)
ON
1)
3.4
– –