Instructions for use Primus Infinity Empowered SW 4.5n 227
Cleaning, disinfection and sterilization
Reprocessing list
Applicable to non-infectious patients.
The list contains approximate values only. The in-
structions of the hospital’s infection control officer
shall prevail and must be observed by the user!
Items which can be reprocessed Recommended reprocessing intervals
1)
With filter at
Y-piece
With microbial
filter at inspira-
tory and expira-
tory port
Without filter
Primus Infinity Empowered workstation Front daily, other sides weekly
Power cable, compressed gas hoses, potential
equalization cable
Monthly
Breathing hoses Daily Per patient Per patient
Y-piece Daily Per patient Per patient
Breathing bag with connector and hose Daily Daily Per patient
Flexible arm for breathing bag (optional) Weekly Weekly Per patient
Ventilator diaphragm
2)
Weekly Weekly Per patient
Cover of breathing system with APL valve Weekly Weekly Per patient
Middle and bottom part of breathing system Weekly Weekly Per patient
Expiratory port/inspiratory port Weekly Weekly Per patient
Absorber and insert Weekly Weekly Per patient
Spirolog/SpiroLife flow sensors Note the instructions for use of the flow sensors.
Anesthetic gas receiving system AGS Note the instructions for use of the AGS.
1) The conditioning intervals depend on the use and position of the filters. The table is merely intended as a rough
guide.
The instructions of the hospital's hygiene officer shall prevail!
2) Drain any water which may have collected in the ventilator diaphragm.
Larger quantities of condensation may impair operation of the workstation and/or lead to failure of the equipment!