Instructions for use Primus Infinity Empowered SW 4.5n 171
Configuration
Setting alarm limits
The following limits may be configured in the menu
alarm limits > alarm limits:
A default alarm limits
B Default limits, anesthetic agents
CAUTION
Risk of inadequate monitoring
Certain monitoring options are mandatory de-
pending on the applicable national requirements.
Some monitoring options may not be covered by
certain screen layout configurations.
Always take national standards into account when
configuring the screen layout.
346
Alarm Setting
ranges
Factory
setting
SpO
2
[%]
81 to 99; – –
1)
80 to 98
– –
92
Pulse
[1/min]
21 to 250
20 to 249
120
50
inO
2
[Vol%]
19 to 99; – –
18 to 98
– –
20
Standby Conf.
ABC
etCO2
[mmHg]
1 to 75
0 to 74
50
– –
inCO
2
[mmHg]
1 to 10 5
MV
[L/min]
0.1 to 20.0
0 to 19.9
12.0
3.0
PAW
[hPa
(cmH2O)]
5 to 99
0 to 35
40
8
1) – –: The corresponding alarm limit is disabled.
Alarm Setting ranges Factory
setting
inHal
[Vol%]
0.1 to 8.4
0 to 8.3
1.5
– –
1)
inIso
[Vol%]
0.1 to 8.4
0 to 8.3
2.3
– –
Alarm Setting
ranges
Factory
setting