Configuration
172
Instructions for use Primus Infinity Empowered SW 4.5n
C alarms in Man.Spont.
The alarms issued when switching to the
Man.Spont. can be activated or deactivated in
Standby. When the alarm limits are set to Yes, the
value set in the automatic ventilation mode is
adopted. For further information, see "Remedying
faults" on page 185.
The following limits may be configured in the menu
alarm limits > misc. alarm settings:
D therapy related
If set to yes, MV alarms and flow apnea moni-
toring are activated in HLM mode.
If set to no, MV alarms and flow apnea monitor-
ing are deactivated in HLM mode.
inEnf
[Vol%]
0.1 to 9.9
0 to 9.8
3.4
– –
inSev
[Vol%]
0 to 9.9
0 to 9.8
4.2
– –
inDes
[Vol%]
0.1 to 21.9
0 to 21.8
12.0
– –
1) – –: The corresponding alarm limit is disabled.
Alarm Factory setting
inO
2 high: yes/no no
MV: yes/no yes
inAgent low: yes/no no
etCO
2: yes/no no
inCO
2: yes/no no
CO
2 apnea alarm cas-
cade: yes/no
yes
378
Alarm Setting ranges Factory
setting
Standby Conf.
DEF
G
Therapy-related alarm limits Factory
setting
MAC low alarm enabled?: yes/
no
yes
alarm limits (MV, Pulse) pre-
sets related to patient weight
and age?: yes/no
yes
MV alarms in HLM?: yes/no yes
CAUTION
Risk of insufficient monitoring
If the setting MV alarms in HLM? is set to no, MV
alarms and flow apnea monitoring are deacti-
vated.
Special attention is required.