Instructions for use Primus Infinity Empowered SW 4.5n 173
Configuration
E device related
– O2 cylinder low alarm active at:
Determine the pressure at which the warn-
ing O
2 CYLIND. LOW is to be issued. This
menu item only appears if the O2 cylinder
has been configured as gas supply,
see page 175.
Factory setting 30 kPa x 100
– soda lime depletion limit *
The absorption capacity of a CO2 absorber
depends largely on the conditions under
which it is operated. Conditions are e.g. flow
levels, gas temperatures, humidity in the ab-
sorber, periods of rest between cases, and
the time of replacement. User practices,
e.g., set fresh-gas flow, ventilation settings,
patient categories entered, and device be-
havior have the biggest influence on these
conditions. Primus Infinity Empowered has
been designed for optimized use of the ab-
sorber.
The total capacity of an unused CLIC ab-
sorber is set by the soda lime depletion
limit. The capacity is entered in absorption
units and can be adjusted if necessary, to
reflect the individual clinical routines de-
scribed above. Decrease setting, if the re-
maining capacity shown is too large and the
alarm does not occur although the soda lime
in the absorber changes color or the inspir-
atory CO
2 concentration inCO2 exceeds the
alarm limit.
Factory setting – –.
– soda lime time limit
Determines the period of use of the soda
lime. As soon as the interval has expired the
advisory message SODA LIME DE-
PLETED? will be issued.
Factory setting 7 days.
– ID ventilation circuit used
Select whether Infinity ID ventilation circuits
are used or not.
When yes is set:
Primus Infinity Empowered automatically
detects the Infinity ID breathing hoses and
checks the following:
- correct hoses used
- assembly complete and hoses connected
to the correct port
- shelf life of hoses.
In case of a deviation a respective alarm
message is issued.
When no is set:
The connected hose system is not checked.
The stored value for compliance is trans-
ferred, see page 124, the value for leakage
is set to "– –".
Factory setting yes.
– Flexible bag arm present?
When yes is set:
Primus Infinity Empowered does not check
for the Infinity ID breathing bag hose.
When no is set:
Primus Infinity Empowered checks for the
Infinity ID breathing bag hose.
Factory setting no
F Other settings
* Only available if the Infinity ID functionality for the
Infinity ID CLIC absorber has been activated by
DrägerService.
Therapy-related alarm limits Factory
setting
open limit menu if alarms oc-
cur?: yes/no
Determine whether or not the
alarm limits menu should appear
automatically when an alarm limit
is violated.
yes
apnea ventilation low priority
only?: yes/no
If set to "no", this enables a cas-
cade for the alarm APNEA VEN-
TILATION, see page 198.
no
immediate alarm upon apnea?
(patients <6 years): yes/no
Further information, see page
172.
no