Instructions for use Primus Infinity Empowered SW 4.5n 99
Operation
Ventilation
Pressure control valve APL
Man.Spont. ventilation mode
On the pressure control valve APL, you can select
between manual ventilation Man. and spontaneous
breathing Spont.
Manual ventilation
z Adjust the APL valve to the desired maximum
airway pressure.
Settings between the stops are also possible.
The patient can be ventilated by hand using the
breathing bag. The pressure is limited to the set
value.
Even in automatic ventilation, the APL valve must
be adjusted to a pressure that is safe for the patient!
Quick release
z Release pressure from the breathing system by
lifting the valve head.
031
WARNING
Risk of patient injury
If the APL valve becomes blocked due to, e.g.,
lines or cables being caught under the valve
head, the patient may be endangered.
Route all cables away from the APL valve. Do
not hang lines, hoses or cables, e.g., the sam-
ple line, on or near the APL valve.
032033
20 3010
20 3010