Cleaning, disinfection and sterilization
234
Instructions for use Primus Infinity Empowered SW 4.5n
3 Fit the water trap at the front.
4 Connect the sample line to the Luer Lock con-
nectors on the Y-piece (A) and the water trap
(B).
5 Make sure the sample line is guided correctly by
using the sample line clip (C). This clip should
be attached to the expiratory port of the breath-
ing system.
Connecting the anesthetic gas
scavenging system AGS
1 Install the particle filter (A), tighten the union
nut.
2 Screw in the flow hose (B).
3 Tighten the union nut (C).
4 Refit the container for buffer volume.
044063
NOTE
Primus Infinity Empowered (no accessories) is
not made with natural rubber latex.
To minimize the risk of exposure to latex, use la-
tex-free breathing bags and breathing hoses.
A
B
C
NOTE
Only use original sample line – other lines may
change the technical data of the device.
055
A
B
C