Instructions for use Primus Infinity Empowered SW 4.5n 45
Operating concept
Loading therapy-related data from
Infinity ID breathing circuits
When an Infinity ID breathing circuit containing
valid therapy-related data is connected to the
breathing system ports of Primus Infinity
Empowered, a field (A) listing all the suggested set-
tings is shown in the Standby mode.* There are
black and gray values displayed in the field. All val-
ues that are suggested ventilation settings for the
transferred ventilation mode (e.g., Volume Mode)
are displayed in black (e.g., V
T). Derived values of
other ventilation parameters are displayed in gray.
The gray values (e.g., P
INSP) do not belong to the
suggested ventilation mode (e.g., Volume Mode).
They are used as suggestions if the ventilation set-
tings have been accepted, but not yet confirmed
and another ventilation mode (e.g., Pressure
Mode) has been selected.
The gray values are also used as suggestions for
ventilation settings when changing the transferred
ventilation mode in Standby (e.g., to Pressure
Mode).
1 To confirm, touch the accept softkey (B) and
press the rotary knob.
After accepting the suggested settings, the values
for the ventilation settings are preconfigured and
can be adapted by the user. When confirming the
settings and ventilation mode with the rotary knob,
the ventilation starts.
The user can reject the suggested parameters from
the Infinity ID breathing circuit.
2 Press the soft key reject (C).
All settings have to be made by the user.
Therapy-related data can only be loaded in
Standby, Monitoring, Man.Spont., and Ext. Out-
let mode. Data for fresh-gas settings are not loaded
in Monitoring, Man.Spont., and Ext. Outlet mode.
During operation of Primus Infinity Empowered
therapy-related data are periodically transferred to
the Infinity ID breathing circuits. The data stored on
the Infinity ID hose set only remain valid for two
hours, if not used.
* Only available when TVS (Transfer of Ventilation Set-
tings) has been enabled by DrägerService and is in the
Standby configuration.
001
A
B
C
WARNING
Risk of patient injury
If incorrect device settings are loaded from
Infinity ID breathing circuits, the patient may
receive the wrong treatment.
The user must always check the settings prior
to confirming any data transfers to ensure the
integrity of the therapy. The user is also re-
sponsible for rejecting or modifying data sug-
gested for transfer prior to activating a venti-
lation mode if the data is inappropriate for the
patient connected.