16 Instructions for use Primus Infinity Empowered SW 4.5n
Application
Intended use
Primus Infinity EmpoweredAnesthesia worksta-
tion for adults, children, and neonates with auto-
matic and manual ventilation, as well as spontane-
ous breathing with or without pressure support.
Utilization
Inhalation anesthesia in rebreathing systems
Inhalation anesthesia in semi-closed to virtually
closed systems with "low flow" and "minimal
flow" techniques (for minimal gas and anes-
thetic agent consumption)
Inhalation anesthesia in non-rebreathing sys-
tems, with a separate fresh-gas outlet for con-
necting, e.g., a Bain system or Magill system,
with a fresh-gas flow of 0.2 to 18 L/min (op-
tional).
Optional:
Operation without nitrous oxide: The device
may only be operated with O
2 and air.
Ventilation modes
Volume-controlled ventilation in Volume Mode.
Optional activation of: Synchronization,
Press. Support (Pressure support) (optional)
Pressure-controlled ventilation in Pressure
Mode
Optional activation of: Synchronization,
Press. Support (Pressure support) (optional)
Manual ventilation (Man.)
Spontaneous breathing (Spont.)
Pressure-supported spontaneous breathing
Press. Support
CPAP (optional)
Volume AF (Volume Mode AutoFlow) (optional)
Optional activation of: Synchronization,
Press. Support (Pressure support) (optional)
The following measured values are dis-
played
Peak pressure PEAK, mean pressure PMEAN,
plateau pressure PLAT, and PEEP
Expiratory minute volume MV,
difference between insp. and exp. minute vol-
ume MVLEAK,
patient compliance C
PAT,
tidal volume V
T,
breathing rate freq.
Inspiratory and expiratory concentration of O
2,
N2O, anesthetic gas, and CO2
O2:
Difference between inspiratory and expiratory
O2 concentration
Optional:
Functional oxygen saturation (SpO
2) and pulse
rate
WARNING
Risk of device failure and/ or danger to patient
If the intended use of this anesthesia machine
is not adhered to, it may fail and/or the patient
may be endangered.
Use the anesthesia machine only as specified
in the intended use of these instructions for
use.