Instructions for use Primus Infinity Empowered SW 4.5n 75
Getting started
Check List
The Check List is displayed after about
35 seconds.
The test steps to be performed are grouped in five
categories:
A Gas Supply
– Pipeline pressure
– Cylinder pressure
– O
2 Flush functional?
– Safety O2 control functional?
B Vaporizers
– Correctly locked in positon?
– Set to zero?
– Fill level OK?
– Safety filler locked?
C Breathing system
– Fully assembled?
– Correctly connected?
– Gas scavenger connected and flow ad-
justed?
– Soda lime OK?
D Miscellaneous
– Water trap fill level OK?
– Suction OK?
– Emergency resuscitator present and
functional?
E Prepare for the self test:
– Set APL valve to MAN.
– Adjust APL valve to 30.
– Seal Y-piece.
– Connect the sample line.
– Close safety O
2 flow control.
z Check the components as instructed in the
Check List on the screen and as described on
the following pages 83 to 94.
F Confirm the change of the soda lime by press-
ing the soft key soda lime changed*.
G Start the self test by pressing the soft key start
self test or the rotary knob.
If the self test has to be interrupted, e.g., for a quick
start in an emergency:
H Press the cancel test key, see "Emergency
start" on page 89.
303
Primus
AB
C
D
E
Check List
F
G
H
* Only without configured CLIC absorber.
WARNING
Risk of device failure and/or patient injury
Canceling the self test may lead to malfunc-
tions; greater attention is required during op-
eration.
Always perform a complete self test, unless
acting in an emergency situation. If canceled
for an emergency, perform a complete self test
as soon as practicable.