Instructions for use Primus Infinity Empowered SW 4.5n 77
Getting started
ing normal operation. The central O
2 supply and
the O2 cylinder cannot both be configured as not
present at the same time.
Open the reserve gas cylinders which have been
configured as present for the self test and then
close them.
O
2 must be connected for the following self test.
O2-Flush
1 Close the Y-piece = plug (A) firmly onto the
cone.
2 Press the O2+ button (B).
3 Breathing bag (C) inflates with an audible flow.
O2 emergency delivery
1 Close the Y-piece = plug (A) firmly onto the
cone.
2 Press the safety knob (D) for the O2 emergency
delivery to unlock it and turn it to set the flow.
3 Breathing bag (C) inflates with an audible flow.
4 Turn safety knob (D) back to its original position
to discontinue O
2 emergency delivery and
press it inwards.
Auxiliary oxygen flow tube (optional)
z Check the auxiliary oxygen flow tube. Adjust the
flow knob (E) and make sure the float moves
freely over the full range of the flow tube.
067
B
D
A
C
A
089
WARNING
Risk of patient injury
If the patient is connected to the auxiliary oxy-
gen outlet without a means of pressure relief,
high pressure will be applied and the patient
endangered.
Do not connect the patient directly to the aux-
iliary oxygen outlet without ensuring a means
of pressure relief.
2
4
6
8
10
E