Maintenance
240
Instructions for use Primus Infinity Empowered SW 4.5n
Preventive maintenance
The following table shows the preventive mainte-
nance intervals:
WARNING
Risk of faulty components
Device failure is possible due to wear or mate-
rial fatigue of the components.
To maintain the proper operation of all compo-
nents, this device must undergo inspection
and preventive maintenance at specified inter-
vals.
WARNING
Risk of electric shock
Before performing any maintenance work, dis-
connect all electrical connectors and gas con-
nectors from power supply and gas supply.
Component Interval Measure Personnel responsible
CO
2 absorber When color changes
(specify color)
Replace Users
AGS filter Replace when blocked. Replace Users
Filter of the endo-
tracheal aspiration
system
Replace every two
weeks.
Replace Users
Upper diaphragm of
ventilator unit
Must be replaced after
one year at most.
Replace Users