Technical data
254
Instructions for use Primus Infinity Empowered SW 4.5n
384
Pressure and flow characteristics of the breathing system without breathing hoses and filters
(in accordance with ISO 8835-2 and ISO 80601-2-13)
pressure [hPa]
Pinsp [hPa] max. Pexp [hPa] max.
flow [L/min]
Minimal Limited Pressure
(in accordance with ISO 8835-5
and ISO 80601-2-13)
–3 hPa
(–3 cmH2O)
Fresh-gas outlet for non-rebreathing systems (optional)
All values apply under STPD conditions:
Connection ∅ 22 mm ISO cone (male) with ∅ 15 mm ISO cone (female)
Pressure limitation Max. 80 hPa (80 cmH
2O) at 18 L per minute
Fresh-gas flow 0 and 0.2 to 18 L per minute
(see fresh-gas delivery for tolerances)
Flowmeter for oxygen delivery – auxiliary oxygen flow tube (optional)
All values apply under STPD conditions:
Connection Graduated connection for use with various hose diameters
Flow range 0 to 10 L per minute
Resolution of the displayed value 0.5 L per minute
Accuracy of the flow display ±10 % of full scale