Instructions for use Primus Infinity Empowered SW 4.5n 253
Technical data
Volume absorber
Reusable absorber canister, filled 1.5 L
Clic absorber
(Drägersorb CLIC 800 Plus) 1.3 L
Clic absorber
(Drägersorb CLIC Free) 1.2 L
Flexible arm for breathing bag
Volume 0.13 L
Compliance 0.13 mL/hPa (0.13 mL/cmH
2O)
Total system leak
(as per ISO 8835-2)
<150 mL per minute at 30 hPa (30 cmH
2O)
Standardized to BTPS conditions.
Pressure limitation valve APL
Setting ranges 5 to 70 hPa (5 to 70 cmH
2O)
Accuracy, within a range of 5 to 15 L
per minute
±15 % of set value or ±3 hPa (±3 cmH
2O),
the greater value applies
Pressure drop at 30 L per minute 2.8 hPa (2.8 cmH
2O) (in Spont position; wet and dry)
Resistance Reusable absorber or CLIC absorber with or without flexible
bag arm, normal operation (filled with Drägersorb 800 +)
Inspiratory Expiratory
As per ISO 8835-2, dry, max. ±6 hPa
(±6 cmH
2O), with hose set for adults
M30146
–4.4 hPa
(–4.4 cmH2O)
4.2 hPa
(4.2 cmH2O)
As per ISO 8835-2, dry, sole breath-
ing system without patient hoses
–3.5 hPa
(–3.5 cmH2O)
3.2 hPa
(3.2 cmH2O)