EasyManua.ls Logo

Dräger Primus Infinity Empowered - Device Check; Power on

Dräger Primus Infinity Empowered
277 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting started
74
Instructions for use Primus Infinity Empowered SW 4.5n
Device check
Prerequisites:
The device has been prepared and assembled
ready for operation (see "Assembly" on page 229).
The gas supply and power supply must be con-
nected.
Power on
1 Switch on Primus Infinity Empowered: press the
system power switch (A), an acous-
tic tone sounds.
After approx. 15 seconds all LEDs and the loud-
speaker are tested by Primus Infinity
Empowered.
The initial screen appears after about 20 seconds.
Primus Infinity Empowered now loads its software
and tests its internal memory.
WARNING
Risk of explosion, fire
If an oxygen leak is suspected within or near
the anesthesia machine, do not initiate opera-
tion.
Disconnect all oxygen supplies and contact a
trained service technician.
CAUTION
Risk of inadvertent movement
If not properly secured, the device may move in-
advertently during operation.
Apply the brakes on the device to ensure it cannot
be moved accidentally.
020
NOTE
If all LEDs do not light up upon initialization, con-
tact DrägerService.
A

Table of Contents

Other manuals for Dräger Primus Infinity Empowered

Related product manuals