Cleaning, disinfection and sterilization
228
Instructions for use Primus Infinity Empowered SW 4.5n
Items which can be re-
processed
Recommended re-
processing inter-
vals
Preclean-
ing
Machine
cleaning
and disin-
fection
Manual Steriliza-
tion
Cleaning Disinfec-
tion
Primus Infinity
Empowered worksta-
tion
Daily no no Outside no no
Power cable, com-
pressed gas hoses, po-
tential equalization ca-
ble
Monthly no no yes no no
Breathing hoses Daily yes yes Observe correspond-
ing instructions for use.
yes
Y-piece no yes
Breathing bag with con-
nector and hose
Daily yes yes Observe correspond-
ing instructions for use.
yes
Flexible arm for breath-
ing bag (optional)
Weekly yes yes no yes yes
Ventilator diaphragm
1)
Weekly no yes no yes yes
Cover of breathing sys-
tem with APL valve
Weekly yes yes no yes yes
2)
Middle and bottom part
of breathing system
Weekly yes yes no yes yes
2)
Expiratory port/inspira-
tory port
Weekly yes yes no yes yes
Absorber and insert Weekly yes yes no yes yes
Spirolog/SpiroLife flow
sensors
Note the instructions for use of the flow sensors.
Anesthetic gas receiv-
ing system AGS
Note the instructions for use of the AGS.
1) Drain any water which may have collected in the ventilator diaphragm.
Larger quantities of condensation may impair operation of the workstation and/or lead to failure of the equipment!
2) The valve plate must be sterilized after washing in order to dry it. Correct operation of the workstation may be impaired and
lead to failure of the workstation if the control areas in the valve plate are not dried completely.