Technical data
266
Instructions for use Primus Infinity Empowered SW 4.5n
Auxiliary power outlets 2 outlets with automatic circuit breakers rated at 3 A each;
1 connection for Desflurane vaporizer with two fuses rated
2 A each and one automatic circuit breaker rated 4 A
Connection for optional halogen light 12 V max. 20 W
Compressed gas supply at workstation pipeline pressure inlet
All values apply under ATPD conditions:
Supply pressure for
O
2, N2O, and Air
2.7 to 6.9 kPa x 100
(39 psi to 100 psi)
Short-term peak inlet flows approx. 55 L/min,
approx. 18 L/min, averaged over 10 s
Scavenging flow for anesthetic gas
receiving system
30 to 50 L per minute
Dew point >5 °C (>41 °F) under ambient temperature
Oil content <0.1 mg/m
3
Particles dust-free air (filtered, with pore size <1 µm)
Driving gas consumption None
Noise emissions Sound pressure level (measured in a free field in accordance
to ISO 3744) at a height of 1.5 m in a distance of 1 m to the
front of the anesthesia system:
Lpmin (Standby) ≤35 dB(A)
Lp
max
("Peak" during ventilation) ≤46 dB(A)
Leq (5 cycles) ≤42 dB(A)