Instructions for use Primus Infinity Empowered SW 4.5n 251
Technical data
tidal volume V
T (system compliance compensated)
All values apply under ATPD conditions:
In Volume Mode and
Volume AF Mode
20 to 1400 mL
1)
With optional Pressure Support 5 to 1400 mL
1)
Accuracy
Within a range of 5 to 150 mL ±10 % of set value or at least ±10 mL, the greater value ap-
plies
Above 150 mL ±5 % of set value or at least ±15 mL, the greater value applies
Frequency 3 to 100 per minute
Accuracy ±10 % of set value or ±1 per minute, the lower value applies
Minimum frequency Freq
MIN
In Pressure Support 3 to 20 per minute or OFF
In Press. Support
CPAP OFF
Accuracy ±10 % of set value or ±1 per minute, the lower value applies
TINSP 0.2 to 6.7 seconds
Insp./Exp. ratio I:E Derived from frequency and T
INSP
Range: max. 5:1 to 1:99
Inspiratory pause T
IP:TINSP 0 to 60 %
Inspiratory flow Derived from V
T and TINSP
In Volume Mode 0.1 to 100 L per minute ±10 %
In Volume AF Mode Max. 150 L per minute +10 %
In Pressure Mode Max. 150 L per minute +10 %
1) Due to gas measurement sampling, leakage (both at the patient and in the device), and resistance/compliance of the pa-
tient and the breathing circuit, the maximum delivered tidal volume may be limited.