Instructions for use Primus Infinity Empowered SW 4.5n 161
Monitoring
Only the DEC-8 or DEC-4 extension lead (purple
plug connector) may be used.
The new sensors are downward-compatible with all
modules already used in the field in older Dräger
machines.
Note the instructions for use of the sensors.
z Select a sensor in accordance with the following
criteria:
– Weight of the patient
– Mobility of the patient
– Possible application point
– Perfusion of the patient
– Duration of use
The following table will assist when selecting a spe-
cific sensor. The table indicates all available sen-
sors and their characteristic values.
Sensor type OxiMax
MAX N
OxiMax
MAX I
OxiMax
MAX P
Dura-
sensor
DS-100 A
OxiMax
MAX A
OxiMax
MAX R
OxiMax
MAX
FAST
Age group Neonates/
Adults
Infants Children Adults
Weight of the
patient
<3 to >40 kg
(<6.6 to
>88 lbs)
1 to 20 kg
(2.2 to
44 lbs)
10 to 50 kg
(22 to
110 lbs)
>40 kg
(>88 lbs)
>30 kg
(>66 lbs)
>50 kg
(>110 lbs)
>40 kg
(>88 lbs)
Duration of
use
Short and long-term monitoring Short-term
monitoring
Short and long-term monitoring
Mobility of
the patient
Limited activity Inactive
patients
only
Limited
activity
Inactive
patients
only, must
be
checked
at least
every
8 hours
Limited
activity
Preferred
measuring
point
Ball of the
foot
Toe Finger Nose Forehead