Instructions for use Primus Infinity Empowered SW 4.5n 177
Configuration
The following parameters can be set in the menu
ventilator and gas delivery > weight related set-
tings > body weight related ventilator settings:
E body weight related ventilator settings
– VT and freq. presetting related to ideal
body weight: (F)
If the settings for V
T and freq. are to be re-
ferred to the patient's body weight, the initial
value for V
T can be selected in accordance
with the Radford nomogram.
Factory setting: yes
– preset configuration: (G)
z Select, edit, and confirm the V
T to be changed
via the rotary knob.
The settings for V
T are interpolated for weights
between the four predetermined classes.
– restore factory default presets (H).
z Select and confirm to restore the factory set-
tings.
The default settings are activated immediately
upon exiting the configuration menu.
z Press exit (I), to exit the menu.
379
Standby Conf.
E
F
G
H
I
weight
(ideal BW)
VT [mL] Freq.
[kg] Setting
ranges
Factory
settings
[1/min]
2 10 to 25 15 35
15 60 to 150 110 26
65 300 to 500 450 13
100 550 to 800 700 10