Instructions for use Primus Infinity Empowered SW 4.5n 205
Remedying faults
Note
(10)
LOW AIR SUPPLY Compressed air supply has
failed.
Open optional Air reserve
gas cylinder.
Check central gas supply,
see page 187.
The central supply is not
connected or the com-
pressed air hose is kinked.
Check connection to cen-
tral gas supply.
Optional air cylinder is
empty or closed.
Connect a full air cylinder
or open the cylinder valve.
Compressed air compres-
sor has failed.
Check compressor.
Note
(10)
LOW N
2O SUPPLY N2O supply has failed. Open N2O reserve gas cyl-
inder.
Check central gas supply,
see page 187.
Probe for the central gas
supply not plugged in or
N
2O hose kinked.
Check connection to cen-
tral gas supply.
N
2O cylinder empty or
closed.
Connect a full N2O cylinder
or open the cylinder valve.
Caution
(11)
LOW O
2 SUPPLY O2 supply has failed. Open O2 reserve gas cylin-
der.
Check central gas supply.
Probe for the central gas
supply not plugged in or O
2
hose kinked.
Check connection to cen-
tral gas supply,
see page 187.
O
2 cylinder empty or
closed.
Connect a full O2 cylinder
or open the cylinder valve.
Caution/
Note
(14/7)
MAC LOW? * The expiratory xMAC value
has fallen below the lower
alarm limit of the automatic
agent alarm.
Check patient condition.
Confirm alarm, if case is
closed.
Check vaporizer fill level.
Check correct position of
vaporizer.
Check for leakages in
breathing system.
Priority Message Cause Remedy