Instructions for use Primus Infinity Empowered SW 4.5n 207
Remedying faults
Warning/
Note
(25/10)
NO AIR DELIVERY * Compressed air supply has
failed.
Open optional Air reserve
gas cylinder.
Check central gas supply,
see page 187.
The central gas supply is
not connected or the com-
pressed air hose is kinked.
Check connection to cen-
tral gas supply.
Optional air cylinder is
empty or closed.
Connect a full air cylinder
or open the cylinder valve.
Compressed air compres-
sor has failed.
Check compressor.
Warning/
Note
(25/10)
NO N
2O DELIVERY*N2O supply has failed. Open N2O reserve gas cyl-
inder.
Check central gas supply,
see page 187.
Probe for the central gas
supply not plugged in or
N
2O hose kinked.
Check connection to cen-
tral gas supply.
N
2O cylinder empty or
closed.
Connect a full N2O cylinder
or open the cylinder valve.
Warning
(31)
NO O
2 DELIVERY O2 supply has failed. Open O2 reserve gas cylin-
der.
Check central gas supply,
see page 187.
Probe for the central gas
supply not plugged in or O
2
hose kinked.
Check connection to cen-
tral gas supply.
O
2 cylinder empty or
closed.
Connect a full O2 cylinder
or open the cylinder valve.
Warning
(31)
NO SpO
2 PULSE No pulse signal detected
with the SpO
2 measure-
ment for approx.
10 seconds.
Check patient's condition!
Check application of SpO2
sensor.
NiBP measurement on the
same arm.
Measure blood pressure on
the other arm.
Priority Message Cause Remedy