Technical data
260
Instructions for use Primus Infinity Empowered SW 4.5n
O2 measurement – paramagnetic (consumption-free)
Measuring range 0 to 100 Vol%
Resolution of the measurement 0.1 Vol%
Resolution of the displayed value
(for insp. O
2, exp. O2)
1Vol%
Accuracy ±(2.5 Vol% +2.5 % rel.)
CO2 measurement (infrared spectrometry)
Measuring range 0 to 13.6 Vol%
0 to 13.6 kPa
0 to 102 mmHg
(at an ambient pressure of 1013 hPa / 760 mmHg)
Resolution of the measurement 1 mmHg
Resolution of the displayed value
(for etCO
2, inCO2)
1 mmHg
Accuracy ±(0.43 Vol% +8 % rel.)
±(3.3 mmHg +8 % rel.)
Anesthetic gas measurement (infrared spectrometry)
All values given in Vol% at 1013 hPa (760 mmHg) ambient pressure
Measuring range anesthetic gases
Halothane 0 to 8.5 Vol%
Isoflurane 0 to 8.5 Vol%
Enflurane 0 to 10 Vol%
Sevoflurane 0 to 10 Vol%
Desflurane
1)
0 to 20 Vol%
Resolution of the measurement 0.1 Vol%
Resolution of the displayed value
(for insp. and exp.) anesthetic gas)
0.1 Vol%
Accuracy
1)
±(0.2 Vol% +15 % rel.)
1) At respiratory rates up to 60 per minute and an insp./exp. ratio of 1:1.