Technical data
262
Instructions for use Primus Infinity Empowered SW 4.5n
Consumption measurement
Fresh-gas consumption per case 0 to 9999 L per gas (O
2, N2O, AIR; only fresh gas applied by
the gas mixer is taken into account)
Accuracy ±10 % or ±1 L, the greater value applies
Resolution 1 L
Total anesthetic gas consumption per
case (liquid)
0 to 3000 mL per agent (Halothane, Isoflurane, Enflurane,
Sevoflurane, Desflurane)
Accuracy Typically ±25 % or ±2 mL, the greater value applies
Resolution 1 mL
Agent consumption due to patient
uptake per case (liquid)
0 to 3000 mL per agent (Halothane, Isoflurane, Enflurane,
Sevoflurane, Desflurane)
Accuracy Typically ±25 % or ±2 mL, the greater value applies
Resolution 1 mL
Soda lime consumption 0 to 1000 L (pure gas CO
2)
Accuracy Typically ±30 % or ±15 L, the greater value applies
Resolution of limit setting 10 L
SpO
2 measurement – light absorption (optional)
The displayed plethysmogram is a relative indicator of the pulse amplitude. Its scale is not absolute and
can only be used to judge the quality of the SpO
2 measurement.
Measuring range SpO
2 1 to 100 %
Resolution of the displayed value 1 %
Accuracy Depending on the sensor model, applies for DS-100 A
Adults
Within a range of 70 to 100 % SpO
2 ±3 %
Neonates
Within a range of 70 to 100 % SpO2 ±4 %