2. Replacement and adjustment of parts
27460
5/5
27460
Colorimeter Section
!
!!
! Replacing the calibration plate
1. Update the calibration plate data.
NOTE
• Register the new calibration plate data to the system.
Bringing up the display
Maintenance → Extension → Maintenance → Various Adjustment → Colorimeter Calibration
(1) Replace the calibration plate stored in the colorimeter unit with new one.
(2) Click F: Functions on the Colorimeter Calibration display, then click Updating the Calibration Plate Data.
(3) Insert the calibration data CD into the drive.
(4) Click Path Setting to configure the CD drive setting of calibration data.
(5) Click OK.
Data of colorimeter and data in the calibration data CD are shown.
(6) When Click [OK] button to register the data. is shown, click OK.
(7) When Are you sure you want to register? is shown, click OK.
(8) When After changing the Calibration Plate, perform the calibration. Would you like to perform the
calibration? is shown, click OK.
Register the data in the CD to the colorimeter.
Data is updated.
(9) When The data has been upgraded. is shown, click OK.
Confirm the registration.
(10) The screen returns to the Colorimeter Calibration display.
2. Check whether updating was performed normally.
NOTE
• Confirm that the colorimeter data and the calibration plate data are the same.
(1) Click F: Functions on the Colorimeter Calibration display, then click Confirming the Colorimeter and
Calibration Plate Data.
(2) Click Path Setting to configure the CD drive setting of calibration data.
(3) Click OK.
Data of colorimeter and data in the calibration data CD are shown.
(4) It matches the Colorimeter Data or It does not match the Colorimeter Data. is shown.
(5) Click Cancel to return to the Colorimeter Calibration display.