4. Troubleshooting
40600
7/7
40600
Attention message: Tablet Replenishment
No. 00610[J]
Empty the Effluent Tank.###
Attention message release
YES or automatic release
• This attention message will be automatically closed when the effluent tank is emptied.
NOTE
• CD, BF or STB
The name of processing solution causing this attention message will be displayed in the second line of the message.
NOTE
• If the Effluent Float Switch turns on, all water supply pumps and cleaning pumps stops physically.
Condition
Check Point
Diagnosis
NOTE
• To access the wiring diagrams, click Wiring diagrams in Bookmarks.
Suffix number Condition
00001 The CD effluent float switch turns on.
00002 The BF effluent float switch turns on.
00004 The STB effluent float switch turns on.
00008 P2 + PS Effluent Float Switch turns on.
1 Check if the float of the effluent float switch moves up and down smoothly.
Failed part Manual No.
Effluent float switch ☞63310
Processor control PCB
☞66200