Processor Standard Setting
1/1
3. Mode
33550
33550
Pump Output Amount Setting [J]
Bringing up the display
Maintenance: 2260 → Extension → Processor Settings → Processor Standard Setting → Pump Output Amount Setting
!
!!
! Explanation
!
!!
! Pump Output Amount Setting (Input range: 0.1 to 200 ml)
Measure the pump output amount in Pump Output Amount Measurement in Functions, and check the measured value.
!
!!
! Functions
!
!!
! Pump Output Amount Measurement
Be sure to measure the output amount when installing it.
If the output amount setting is not appropriate, the water supply or replenishment operation is not carried out properly.
!
!!
! Initial Replenishment Operation
The service personnel password (2260) is required to display this.
After operating each replenisher pump, water supply pump, refilling water pump for a specified time, the air remaining in the hose can
be extracted.
To cancel the initial replenishment, click Cancel.
IMPORTANT
• Be sure to carry out the initial replenishment operation at system installation.
• When the initial replenishment operation has been carried out if air in the hose cannot be removed, activate the pump
of the relevant hose to remove air using the pump output amount setting of Functions.
M5453-00
Distributed by: minilablaser.com