4. Troubleshooting
45700
2/4
45700
Error message: SM replenishment
No. 05708[SM]
Replenishment Package solution remaining error.
No. 05709[SM]
Replenishment Package solution remaining error.
Error message release
YES
Condition
The message No. 0700 Replace the Replenishment Package with a new one. is shown when the replenishment alarm value is set to -5%
(initial value) in Package Capacity Setting and either replenishment package becomes empty. In this case, the message Nos. 5708 and
5709 Replenishment Package solution remaining error. is not shown.
Check Point
Diagnosis
NOTE
• To access the wiring diagrams, click Wiring diagrams in Bookmarks.
Suffix number Condition
00000 The percentage of the replenishment remaining amount (internal calculated value) to the replenishment package
capacity became smaller than the replenishment alarm value. (Replenishment remaining amount < Package
capacity × Replenishment remaining amount / 100) But the replenishment solution sensor does not turn DARK.
1 Check if the actual package capacity is same as the one set in the Package Capacity Setting.
2 Check if the replenishment alarm value is sufficient.
3 Check that the valve of the strainer unit is not closed.
4 Clean the strainer.
5 Check that the replenisher hose and water supply hose are not clogged.
Adjustment failure point Manual No.
Sensitivity adjustment failure of the replenishment solution sensor ☞27830
Failed part Manual No.
Replenishment solution sensor CD-A, CD-B, CD-C, BF-A, BF-B, or STB
☞63300
SM I/O PCB
☞66600
Processor control PCB
☞66200
Distributed by: minilablaser.com