3/16
4. Troubleshooting
45900
45900
Error message: F replenishment
No. 05901[F]
Replenishment solution level sensor error. P1R
No. 05902[F]
Replenishment solution level sensor error. P2RA
No. 05903[F]
Replenishment solution level sensor error. P2RB
No. 05904[F]
Replenishment solution level sensor error. PSR
Error message release
YES
Condition
Diagnosis
NOTE
• To access the wiring diagrams, click Wiring diagrams in Bookmarks.
Suffix number Condition
00000 When # # # (upper) replenishment solution level sensors of each processing solution are ON and when # # #
(lower) replenishment solution level sensors are OFF
Countermeasure
1. Check the status of the replenishment solution level sensors of each processing solution with Input Check.
☞35220
2. Perform recovery procedure 2 from the replenishment error.☞4620
Failed part Manual No.
Replenishment solution level sensor ☞63280
F replenishment I/O PCB ☞66800
Processor control PCB
☞66200
Distributed by: minilablaser.com