EasyManuals Logo

GE LOGIQ P5 Service Manual

GE LOGIQ P5
532 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
GE HEALTHCARE
DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
5-24 Section 5-4 - Main Board Detail
5-4-4-2 BIOS Content (cont’d)
Figure 5-22 Boot menu in BIOS
Figure 5-23 Security Menu in BIOS
Main Advanced Boot Securit
y
Power Exit
Boot Device Priority
Boot Priority Selection [Type Based]
1
ST
Boot Device [USB CDROM]
2
n
d
Boot Device [Primary Master]
3
r
d
Boot Device [Disabled]
4
t
h
Boot Device [Disabled]
5
t
h
Boot Device [Disabled]
6
t
h
Boot Device [Disabled]
7
t
h
Boot Device [Disabled]
8
t
h
Boot Device [Disabled]
Boot Settings Configuration
Quick Boot [Enabled]
Quiet Boot [Enabled]
Boot Display [Clear]
Automatic Boot List Retry [Disabled]
AddOn ROM Display Mode [Keep Current]
Halt On Error [Disabled]
Ht "Del" Message Display [Disabled]
Interrupt 19 Capture [Disabled]
PXE Boot to LAN [Disabled]
Power Loss Control [Remain off]
BIOS SET UP UTILITY
Main Advanced Boot Securit
y
Power Exit
Security Seetings
Supervisor Password : Installed
User Password : not Installed
Change Supervisor Password
User Access Level [No Access]
Change User Password
Password Check [Setup]
Boot Sector Virus Protection [Disabled]
Hard Disk Security
Hard Disk Security User Passwords
Hard Disk Security Master Passwords
BIOS Update & Write Protection [Disabled]
END-KEY Loads CMOS Default [Yes]
BIOS SET UP UTILITY

Table of Contents

Other manuals for GE LOGIQ P5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P5 and is the answer not in the manual?

GE LOGIQ P5 Specifications

General IconGeneral
Power Supply100-240VAC, 50/60Hz
Display15" LCD
Doppler ModesColor Doppler, Power Doppler
Imaging ModesB-Mode, M-Mode
ProbesConvex, Linear, Phased Array
ApplicationsAbdominal, Vascular, Cardiac, OB/GYN, Small Parts, Musculoskeletal
ConnectivityDICOM, USB, Ethernet
Probe Connectors3
PortabilityMobile
Operating Temperature10°C to 40°C
Relative Humidity30%