DIRECTION 5394141, REVISION 5 LOGIQ™ P5 SEVICE MANUAL
8-154 Section 8-4 - Software Installation
8-4-4-4 Functional Checks for Service Platform (Diagnostic Program) (cont’d)
4.) Select
GE Service at the “Select User Level” field.
5.) Enter the password for the Service Platform.
6.) Click on
Okay.
7.) Verify that the following screen (Service Platform) is displayed on the monitor.
8.) Click on
x located at the upper right corner of the service platform screen to close the Service
Figure 8-211 Select User Level
NOTICE
When the service platform is NOT displayed, check if CAPS lock is selected. The CAPS should not be
selected.