Configure Settings
NOTE:
You can secure some communication channels with encryption and
authentication. GE Healthcare recommends that you use the encrypted
channels, not the unencrypted channels. For example:
• Upgrade the MUSE server 8.0/9.0 API3 version to the MUSE NX API3 version,
to have a secure communication between the MAC 5 system and the MUSE
server.
Configure a DCP Server Destination to Transmit Reports
1. Select Settings > Workflow > Transmission.
2. Select the Add icon to add transmission modes.
3. Select DCP to configure a DCP server destination.
The report is sent to the server using the DCAR Communication Protocol (DCP).
The MUSE server and EMR Gateway use DCP.
• If you configure the DCP server destination to the MUSE system, a Hilltop
format report is sent to the server.
• If you configure the DCP server destination to the an EMR Gateway, a
Sapphire XML and PDF report are sent to the server.
4. Configure the fields in the table to add a DCP server destination.
168  MAC
â„¢
5 A4/MAC
â„¢
5 A5/MAC
â„¢
5 Lite 5864335-001-1