Configure Settings
NOTE:
You can secure some communication channels with encryption and
authentication. GE Healthcare recommends that you use the encrypted
channels, not the unencrypted channels. For example:
Upgrade the MUSE server 8.0/9.0 API3 version to the MUSE NX API3 version,
to have a secure communication between the MAC 5 system and the MUSE
server.
Configure a DCP Server Destination to Transmit Reports
1. Select Settings > Workflow > Transmission.
2. Select the Add icon to add transmission modes.
3. Select DCP to configure a DCP server destination.
The report is sent to the server using the DCAR Communication Protocol (DCP).
The MUSE server and EMR Gateway use DCP.
If you configure the DCP server destination to the MUSE system, a Hilltop
format report is sent to the server.
If you configure the DCP server destination to the an EMR Gateway, a
Sapphire XML and PDF report are sent to the server.
4. Configure the fields in the table to add a DCP server destination.
168 MAC
5 A4/MAC
5 A5/MAC
5 Lite 5864335-001-1